Views: 0 Author: Site Editor Publish Time: 2025-12-22 Origin: Site
During daily use, dosing pump may malfunction due to drug impurities, wear, or improper operation. Below are 8 common faults with concise descriptions and targeted solutions for quick troubleshooting and maintenance.
[Fault Cause]: Blockage from impurities, precipitates, or residual drug accumulation on filter screen/pipeline wall. High-concentration drugs or suspended particles increase blockage risk.
[Fault Phenomenon]: Significantly reduced/no drug flow; negative pressure at inlet or increased equipment load.
[Solution]: Stop equipment, close inlet valve. Disassemble and clean filter screen and pipeline of the dosing pump with water. Reassemble in order, check seals, and test run to confirm fault elimination.
[Fault Cause]: Foreign objects (impurities, crystals) get stuck between the ball valve core and seat, blocking drug flow.
[Fault Phenomenon]: No or fluctuating drug flow; stuck ball valve during manual adjustment.
[Solution]: Stop the equipment, close the inlet/outlet valves. Disassemble the ball valve, clean the core and seat. Check for wear, replace if necessary. Reassemble, tighten, and test run.
[Fault Cause]: Wear on the valve core-seat sealing surface, loose installation, or impurity adhesion affects sealing.
[Fault Phenomenon]: Air leakage at interfaces, unstable outlet pressure, or potential drug back-suction.
[Solution]: Stop the equipment, clean the valve interface. Tighten the ball valve clockwise. Apply sealant if needed. Test run; replace the valve if leakage persists.
[Fault Cause]: Equipment vibration, incomplete installation, or aging causes seal ring loosening/falling off.
[Fault Phenomenon]: Drug leakage at interfaces, insufficient flow, or unstable pressure.
[Solution]: Stop the equipment, close valves. Check 4 seal rings (2 inlet, 2 outlet). Re-embed or replace aging/lost rings. Reassemble, tighten, and test run.
[Fault Cause]: Diaphragm damage due to long-term pressure, wear, chemical corrosion, or installation scratches.
[Fault Phenomenon]: Sharp drop in outlet pressure, minimal flow, or internal drug leakage.
[Solution]: Stop the dosing pump, cut off power, close valves. Disassemble the pump head, replace the damaged diaphragm with a matching new one. Reassemble evenly, test run.
[Fault Cause]: Lack of maintenance leads to rust, oil exhaustion, or foreign objects causing connecting rod seizure.
[Fault Phenomenon]: Failure to start or immediate shutdown; abnormal noise; stuck motor shaft.
[Solution]: Two options based on damage:
Option 1: Maintenance. Disassemble the connecting rod group, grind rusted parts, replace worn components, lubricate, and reassemble.
Option 2: Replacement. Install a new matching connecting rod group if damage is severe, then test run.
[Fault Cause]: Motor burnout due to overload, water ingress/short circuit, or bearing damage leading to rotor seizure.
[Fault Phenomenon]: No response on startup; abnormal temperature rise and burning smell; short/open circuit in windings.
[Solution]: Cut off power of the dosing pump, replace with a matching new motor. Correctly wire, check rotation direction, and test run.
[Fault Cause]: Misoperation or impact shifts the adjustment cam to "0", resulting in zero pump stroke and no drug delivery.
[Fault Phenomenon]: Normal motor operation but no drug output; cam scale shows "0".
[Solution]: Rotate the adjustment cam counterclockwise away from "0" to the required scale. Test run and lock the cam.
content is empty!